Medical Gas Plants

Medical Gas Manifold Plant
HTM 02-01 & EN ISO 7396-1

Medical Gas Manifold Systems

Integral Medical Gas Plants are engineered to provide regular and reliable 24/7 gas service to critical hospital departments such as operating rooms, intensive care units, patient rooms, and emergency services.

The system reduces the pressure of high-pressure gases in cylinders (approx. 200 Bar) such as Oxygen, Nitrous-Oxide, or Carbon Dioxide to the hospital line pressure level (4-5 Bar). Through the Pneumatic Auto Change-Over Panel located between two redundant cylinder groups (Ramp A and Ramp B), the system automatically switches to the full cylinder group without experiencing even a momentary interruption.

Engineering & Safety Advantages

  • Pneumatic Auto Changeover: Electricity-independent switching ensuring zero outage during power cuts.
  • Dual Stage Regulation: High-flow 1st and 2nd stage brass regulators providing smooth line pressure.
  • Emergency Bypass Line: Integrated bypass valves allowing maintenance without interrupting supply.
  • BMS Connectivity: ModBus (RS-485) communication transmitting pressure status to hospital BMS.

Technical Documents & Drawings

PDF Datasheet

Tender Specification Text

Medical gas manifolds shall be manufactured in compliance with HTM 02-01 and EN ISO 7396-1 standards, featuring dual cylinder banks, electricity-independent fully pneumatic auto change-over panel, dual-stage pressure regulators, and BMS (modBus) connectivity.

Technical Parameters & Operating Values
Parameter Value / Standard
Maximum Inlet Pressure 200 - 230 Bar
Outlet Line Pressure 4 Bar - 5 Bar (Adjustable)
Changeover System Fully Pneumatic Auto Change-Over
Regulator Type High-flow 1st & 2nd Stage Brass Regulators
Safety Equipment High-pressure relief valve & bypass line
Standards EN ISO 7396-1, HTM 02-01, MDD 93/42/EEC
Capacity & Part Numbers
Capacity Oxygen (O2) Nitrous Oxide (N2O) Carbon Dioxide (CO2)
2x5 Cylinders 10-005 20-005 11-005
2x10 Cylinders 10-010 20-010 11-010
2x15 Cylinders 10-015 20-015 11-015
2x20 Cylinders 10-020 20-020 11-020
Frequently Asked Questions (FAQ)

The system consists of two cylinder banks (Ramp A and B). When the pressure in one bank falls below the set limit, the pneumatic changeover valve mechanically switches to the other full bank without any interruption or pressure fluctuation in the gas flow to the hospital.

All our gas plants are designed and certified in full compliance with HTM 02-01 (UK Department of Health norms), EN ISO 7396-1, and the Medical Device Directive (MDD 93/42/EEC).

Yes. Our plants feature electronic control panels with the modBus (RS485) communication protocol. Cylinder ramp pressure statuses, changeover alarms, and fault notifications can be transmitted instantly to the hospital's Building Management System (BMS).